HPLC Analytical Conditions
Classification | Condition | Explanation | ||
---|---|---|---|---|
1st condition | Instrument | Agilent 1260 series | ||
Column | Phenyl-Hexyl column | |||
Oven temperature | 25℃ | |||
Wave length | UV 254 nm | |||
Injection volume | 20 μL | |||
Mobile phase | A: 0.1% Trifluoroacetic acid in H2O | |||
B: Acetonitrile | ||||
Time (min) | A (%) | B (%) | ||
0 | 92 | 8 | ||
15 | 92 | 8 | ||
20 | 90 | 10 | ||
25 | 90 | 10 | ||
80 | 53 | 47 | ||
85 | 40 | 60 | ||
90 | 40 | 60 | ||
2nd condition | Instrument | Agilent 1260 series | ||
Column | YMC-Triat C18 (5 μm, 4.6×250 mm) | |||
Oven temperature | 25℃ | |||
Wave length | UV 210 nm | |||
Injection volume | 20 μL | |||
Mobile phase | A : 0.1% H3PO4 in H2O | |||
B : Acetonitrile | ||||
Time (min) | A (%) | B (%) | ||
0 | 95 | 5 | ||
35 | 95 | 5 | ||
40 | 85 | 15 | ||
50 | 85 | 15 | ||
70 | 55 | 45 | ||
75 | 0 | 100 | ||
90 | 0 | 100 |